medical pump
Features:
Large LCD screen, all data synchronized displayed, easy to operate
Long-life, no need for lubricant, accurate control
Integrated ultra-sound bubble detector, accurate and reliable
High quality mini-pressure block detector, accurate and reliable
Auto-controlled multiple KVO function, prevent IV block effectively
Infusion rate can be displayed by "mL/h" or "drop/min", flexible
5 type of infusion sets can be applied, diverse and flexible
Solid appearance design, easy to clean and sterilize
Technical Data
Power consumption: ≤30VA
Dimensions(LWH): 230mm110mm240mm
Net weight: ≤3.5kg(without package)
Electrical standard: Class I
Electrical type: Type BF
Power requirement: Main power:AC100-240V, 50/60Hz
Internal power: NiMH battery 12V(fully charged, ≥4hrs at 25mL/h)
Performance Parameter
Recommended infusion sets: standard silica gel infusion sets, 20 drop/ml
Flow rate setting scale: 1.0 1200mL/h in 0.1mL/h increment
BOLUS rate: 600/ 800/ 1000/ 1200/1500mL/h
PRIME rate: 600/ 800/ 1000/ 1200/1500mL/h
KVO rate setting scale: off, 1.0- 5.0mL/h (in 0.1mL/h increment, auto control)
MAX volume limit: 9999mL(MIN increment 0.1mL/h)
MAX volume accumulated: 9999mL(MIN increment 0.1mL/h)
Accuracy: ±5%
Occluded alarm threshold: 5 level (300/ 500/ 700/ 900/ 1100mmHg)
Bubble detective sensitivity: single bubble volume>=30×10-3mL
Alarm: Failure, Low Battery, AC Failure, Door Open, Occlusion, Air in line, Infusion Finished (auto switch to KVO)
Environment requirement
Operating environment:
Temp: +15 +40, Humidity: 20%-90%, Atmosphere pressure: 70-106kPa
Attitude: ≤2000m
Transportation and storage:
Temp: -20 +55, Humidity:<=93%,,Atmosphere pressure: 50-106kPa
medical pump