Syringe Pump/medical pump

Syringe Pump/medical pump
Product Detailed
Related Categories:Surgical Instruments
The pump can be flexibly used both for patient bed and drip stand Automatically identify the scale of the injector

Features:

1.The pump can be flexibly used both for patient bed and drip stand

2.Automatically identify the scale of the injector

3.Operated by button and knob, convenient and practical

4.LCD screen, clearly to be read

5.Against power supply fluctuation; no injection remaining

Technical Data

Power consumption: ≤30VA

Dimensions(L*W*H): 302mm*240mm*124mm

Net weight: ≤2.8kg(without accessories)

Electrical standard: Class II

Electrical type: Type BF

Power requirement:

 Main power: AC 100-240V, 50Hz/60Hz

 Internal power: NiMH battery 12V (fully charged, ≥4hrs at 5.0 mL/h)

Performance Parameter

Injector: 20ml, 50ml  identified automatically

Injection rate: 0.1mL/h-400mL/h(for 20 mL injector), 0.1mL/h-600mL/h (for 50 mL injector)

BOLUS rate: 400mL/h(for 20 mL injector), 600mL/h(for 50 mL injector)

PRIME rate: 400mL/h(for 20 mL injector), 600mL/h(for 50 mL injector)

KVO: OFF, 0.5mL/h(Injection rate≥5.0mL/h)

Maximal volume limit: 999.9mL(in 0.1mL/h increment)

Maximal volume accumulated: 999.9mL(in 0.1mL/h increment)

Injection tolerance: ±3%

Block alarm: high, middle, low

Monitoring data: Failure, Low battery, AC failure, Block, injection finished, injector empty, Injector dropped, injection near to empty

Memory function: automatically keep a record of the last injection data

Silence function: 2 minutes

Environment requirement

Operating environment:

Temp: +5 ~+40°C,Humidity: 80%,Atmosphere pressure: 70 ~ 106kPa

Transportation and storage:

Humidity:≤93%,Atmosphere pressure: 50 ~ 106kPa