- Contact Person : Ms. Jia Kathy
- Company Name : Beijing Aeonmed Co., Ltd.
- Tel : 86-10-83681616-8296
- Fax : 86-10-63718989
- Address : Beijing,Beijing,No.4 Hangfeng Road, Fengtai Science Park, Fengtai District, Beijing,China.
- Country/Region : China
- Zip : 100070
anaesthetic ventilator machine integrated breath system
Aeonmed is professional anesthesia machine, ventilator manufacturer, medical pump, operating light and table manufacturer since 2001, after more than 10 years development, Aeonmed has been grow up to OR&ICU operating expert.
With owning more than 800 staffs and covering 10,000 square meters manufacturing base, Aeonmed is moving toward to be most famous hospital medical solution expert to assist hospitals fine best solution.
Our main product lines:
Anaesthesia machineTransport and ICU ventilatorInfusion pump and Syringe pumpOperating table and lamp/lightCeiling pandentMedical solution, Hospital information and area gas systemAeonmed is welcome your connection and communication, we believe we are the best choice for your medical solution.
Features. Classical multifunctional anaesthesia machine
. Dynamic compensation system for tidal volume
. Volume & pressure ventilation, SIMV mode with flow trigger
. CE marked, meet clinical safety requirement of EU
. Advanced electromagnetic valve ensures accurate tidal volume and compensation
. Waveforms displayed include: P-T, F-T and V-T
. Solid coating and anti-dust smooth surface
Technical SpecificationPhysical & Environmental Specification. Dimensions (WxHxD): 824x1382x680mm
. Weight: 110kg (including two vaporizers)
. Storage temperature: -22~+55
. Working temperature: +10~+40
. Storage humidity: ≤93%
Electrical Specification. Mains: 100-240VA C, 50-60Hz
. Battery: 24VDC, minimum90min
. Power consumption: ≤50W
Pneumatic Specification. Gas supply: O2, N2O: 0.28-0.6Mpa
. Flowmeters : O2: 0-1L/min; 1-10l/min
N2O:0-1L/min; 1-10L/min
. Gas supply: O2 deficiency alarm
Hypoxic guard system
O2 flush: 35-75L/min
. Working mode: Closed, Semi-closed, Semi-open
. Driven mode: Pneumatically driven and electronically controlled
. Operating mode: Man. / Vent.
. Safety valve: ≤8kpa
Ventilator Specification. Patient type: Adult, pediatric
. Ventilation modes: IPPV, PCV, SIMV(flow-trigger), Manual
. Setting: Touch key & navigator wheel knob
. Tidal volume: 0-1600ml
. Ventilation frequency: 4-100bmp
. I:E ratio: 2:1-1:8
. Pressure control: 5-60cmH2O
. Ppeep: 0-20cmH2O
. Inspiratory plateau: 0-50%Ti
. Trigger sensitivity: 2-30L/min
. High pressure: 21-80cmH2O
. Low pressure: 0-20cmH2O
. Monitoring: Display: 5.7TFT color screen
Pressure: Paw, Ppeak, Pplat, Pmean
Volume: MV, Vt
Breath rate
Gas: FiO2
Compliance
Graph display: P-T, F-T, V-T waveforms
. Alarm: Pressure: Paw, high/low limit
Volumes: MV high/low limit, Vt high limit
Low O2 supply pressure
Apnea
Self-testing failure
Power failure
Alarm silence: ≤120s
Vaporizer. Support 2 vaporizer (Selectatec compatible)
. Agent type: Halothane, Enflurane, Isoflurane, Sevoflulrane available.
anaesthetic ventilator machine integrated breath system